Digital Imaging for Optimal Screening:
Diabetes, Glaucoma Macular Degeneration
and other Retinal Disease |
|
|
|
| |
The Nidek Non-Mydriatic NM-1000 Fundus Camera
provides high-resolution megapixel 45-degree images by incorporating a
new Digital Progressive Scan CCD Camera. Unlike conventional fundus cameras
that convert digital signal into analog data for storage, the Nidek NM-1000
Fundus Camera offers direct, digital-to-digital data management to provide
images of the highest quality.
|
|
 |
|
| |
|
|
| |
|
|
|
|
| |
Large LCD provides easy operation |
|
|
|
| |
The large 6.4-inch LCD specifically designed to position the control
panel close to the monitor enables easy operation of the NM-1000.
In alignment with infrared light, the internal fixation target, patient
data, photographed data and other setup information are displayed for
maximum convenience.
|
|

|
|
| |
|
|
| |
|
|
|
|
| |
Excellent Functionality |
|
|
|
| |
Photographed images are immediately displayed on the monitor in color,
where static imaging and segmental zooming are possible.
The large 6.4-inch color LCD and control panel allow greater ease of operation
and functionality for image capture, processing, editing and transmission
of image data to the removable ATA flash card.
|
|

|
|
| |
|
|
| |
|
|
|
The data files can be temporarily stored in a PC card (ATA flash card).
Data storage capacity will be dependent upon the PC card memory.
|
|
 |
| |
|
|
|
|
| |
Total Data Management |
|
|
|
| |
The NM-1000 streamlines your data, including
signals from the diagnostic and photographic CCD camera, and information
from all other NM-1000 functions.
The NM-1000 Camera head swivels 15 degrees to
the left or right for maximum flexibility.
|
|
 |
|
| |
|
|
| |
|
|
|
|
| |
NAVIS Compatible |
|
|
|
| |
|
|
|
|
| |
USB interface provides faster transmission of high-resolution images
to a computer.
The Nidek NM-1000 Fundus Camera is the first
camera in the market to incorporate USB interface.
Using this interface, the system can be integrated with Nidek
Advanced Vision Information System (NAVIS)
for practical data management.
With NAVIS, storage, retrieval and management of diagnostic images is
easily accomplished.
|
|
 |
|
| |
|
|
|
|
| |
Patient Comfort |
|
|
|
| |
Various image modifications such as image processing, analysis, editing
and transmission, are also possible. The built-in digital CCD camera requires
lower flash illumination as compared to film photographs, so the patient
will experience minimal discomfort during the photography session.
|
|
 |
|
| |
|
|
| |
|
|
|
|
| |
Highly Flexible |
|
|
|
| |
|
|
|
|
| |
USB, RGB, Analog and NTSC signal outputs are incorporated into the design,
allowing quick and easy connection to other instruments.
USB interface enables connection to a PC workstation and a NAVIS
network.Three independent ports (R/G/B) allow single-color combination
images to be displayed.
NTSC interface allows images stored in the camera unit to be displayed
on a TV monitor.
|
|
 |
|
| |
|
|
|
|
| |
System Structure |
|
| |
|
|
|
|
| |
 |
- Head rest
- Front lens
- Optical head
- Control panel
- Joystick
- Chin rest adjustment
|
|
|
| |
|
|
|
|
| |
NM-1000 Specifications |
|
| |
|
|
|
|
| |
|
Main Camera Body
|
|
|
Type
|
Non-Mydriatic Fundus Camera
|
|
Picture angle
|
45 degrees
|
|
Working distance
|
43 mm (from camera to cornea)
|
|
Working distance detection method
|
Purkinje bright spots
|
|
Minimum diameter of pupil
|
¢ 4 mm
|
|
Dioptric compensation for patient's eyes
|
-32 D ~ +40 D
-32 D ~ -10 D with "-" compensator
-10 D ~ +14 D without compensator
+14 D ~ +40 D with "+" compensator
|
|
Focusing method
|
Split bright target
|
|
Illumination for observation
(with an infrared filter)
|
Halogen lamp; limitless intensity setting
|
|
Flash for photographing
|
Xenon lamp; 8 intensity setting
|
|
TV camera for observation
|
Built-in 1/3 inch CCD (B&W)
|
|
TV camera for photographing
|
Built-in 1/2 inch digital progressive scan CCD
|
|
TV monitor
|
6.4 inch LCD color
|
|
Flash recycle time
|
5 sec. repeatable
|
|
Internal fixation targets
|
Manual lever
|
|
External fixation targets
|
Free-arm (option)
|
|
Data imprinting
|
Patient data (ID, date)
Photographic data (flash intensity)
|
|
Saving temporary images function
|
Compact flash card (PC card adapter: PCMCIA Type II)
|
| |
|
|
Camera Operation Stand
|
|
|
Horizontal movement
|
65 mm forward/backward
(Fine movement: 10 mm forward/backward and left/right)
|
|
Vertical movement
|
30 mm
|
|
Function of joystick
|
horizontal, vertical movement, shutter release
|
|
Signal outlet
|
USB (Ver. 1.1)
|
|
Video outlet
|
RGB / NTSC composite
|
|
Power supply
|
AC 100/120/230 V ± 10%, 50/60 Hz
|
| |
|
|
Chin Rest
|
|
|
Vertical movement of chin rest
|
65 mm
|
|
External fixation targets
|
Free-arm (2 joints - option)
|
|
Dimension / Weight
|
360 (W) x 530 (D) x 565 (H) mm / 26 Kg
13.8 (W) x 20.5 (D) x 22 (H) inch / 57 Lbs
|
|
|
| |
|
|
|
|